Cureus | Electrocardiographic Interference on Bispectral Index Monitor: A Case of Crossed Wires

2022-09-25 16:25:32 By : Mr. Kent Wong

"Never doubt that a small group of thoughtful, committed citizens can change the world. Indeed, it is the only thing that ever has."

Cureus is on a mission to change the long-standing paradigm of medical publishing, where submitting research can be costly, complex and time-consuming.

The SIQ for this article will be revealed

electrocardiogram (ecg/ekg), electrocardiogram, monitoring interference, neuro-monitoring, bis, bispectral index monitoring

Alexander S. Doyal , Hannah E. Doyal, David N. Flynn, Jay Schoenherr, Jane J. Kwon, Priya Kumar

Cite this article as: Doyal A S, Doyal H E, Flynn D N, et al. (September 23, 2022) Electrocardiographic Interference on Bispectral Index Monitor: A Case of Crossed Wires. Cureus 14(9): e29511. doi:10.7759/cureus.29511

The Bispectral Index (BIS) has been widely utilized to monitor patients’ levels of consciousness during anesthesia. Despite its practicality and prevalence, BIS monitors have been reported to show erroneous readings due to various factors that interfere with the proper reading of the brain’s electrical activity. We present a case where the BIS monitor misinterpreted the patient’s cardiac activity as her neural activity and resulted in a falsely elevated BIS number despite proper placement and lack of underlying patient medical condition, including neurological injury. It is crucial to remain vigilant about monitoring and understanding BIS readings to assess patients’ awareness and effectiveness of anesthesia properly.

The Bispectral Index (BIS, Aspect Medical Systems, Newton, MA, USA), monitors the patient’s neural response to the hypnotic effect of general anesthesia through a proprietary interpretation of the raw electroencephalogram (EEG). Maintenance of general anesthesia is typically titrated to maintain a BIS numerical rating between the range of 40 and 60. Many different conditions may influence the BIS monitor, leading it to reveal an incorrect state of hypnosis. We will present our case report and then describe various interferences described in the literature.

A 46-year-old woman (consent acquired) with a past medical history only significant for morbid obesity (body mass index 43), underwent a laparoscopic gastrectomy for weight loss. She was premedicated in the preoperative holding area with 2 milligrams (mg) of midazolam and was then transferred to the operating room. She was pre-oxygenated with 100% oxygen. Anesthesia was induced with 100 micrograms (mcg) fentanyl, 100 mg lidocaine, 200 mg propofol, 16 mg cisatracurium, and 20 mcg dexmedetomidine. She was intubated with a 7.0 endotracheal tube by direct laryngoscopy using a mac 3 laryngoscope. General anesthesia was maintained with sevoflurane and cisatracurium. Higher levels of sevoflurane were used to deepen the anesthetic secondary to the patient’s movement (end-tidal concentration ranging from 1.8% to 3.7%). 

The surgery commenced without incidence. A BIS monitor was placed in a typical manner and adequate SQI signals were obtained. At first, the BIS values were in the 40-60 range and were as expected for the patient's depth of anesthesia. However, later on in the surgery, it was noted that the BIS monitor intermittently showed a waveform that mirrored the patient’s electrocardiogram (EKG) activity. This occurred while the end-tidal sevoflurane concentration was particularly high at 3.7%. Under such conditions, the BIS monitor would generally be expected to be low. The clinicians observed that the displayed BIS number appeared to be influenced by the electrical interference of the patient’s EKG (Figures 1, 2).

The surgery proceeded successfully and without complications, BIS values were in the 40-60. At the conclusion of the surgery, the patient was awakened and extubated without incident. The BIS values during this time frame appropriately increased during emergence, progressively climbing into the 80s. She was then transferred to the post-anesthesia care unit (PACU). Her time in the PACU was unremarkable, and she was transferred to the floor. She was discharged home on the following day.

The BIS may be affected by a variety of physiological states of the patient. These include a diverse range of various physiological conditions such as hypoglycemia, hypothermia, cerebral ischemia, the post-ictal state, and low voltage EEG [1]. Pharmacological intraoperative exposure to medications such as ketamine or nitrous oxide will also lead to erroneous BIS readings [1].

The BIS monitor may also be affected by external influences. The BIS constantly evaluates the electrical activity of the brain by interpreting the faint signals of the EEG. Because the monitor is in essence observing small electrical waveforms, it may be susceptible to external electromagnetic interference. The operating room is a rather electrically noisy environment. All electrical equipment produces a field of electromagnetic radiation to varying degrees.

Several external sources that lead to erroneous data have been identified and previously reported. These include electromagnetic monitoring systems, surgical endoscopic shavers, warming blankets positioned on the patient’s head, noise from cardiopulmonary bypass machines, a patient in deep hypothermic cardiac arrest with isoelectric EEG [2], and electromyographic monitoring endotracheal tubes [3]. These sources of interference should be filtered out and ignored as noise by the BIS computational unit. However, this interference may also be interpreted as the patient’s actual EEG signals, leading to an erroneously high BIS number [1].

The appearance of the EKG tracing on the BIS monitor has been previously reported in patients with severe neurological injury resulting in burst suppression and an isoelectric EEG in the intensive care unit [4,5]. In the absence of neurological electrical activity, the BIS erroneously reported cardiac electrical activity as cortical in origin. This could have led to the providers thinking that a brain-dead patient continued to have some neurological function. In another example, a display of cardiac activity by a BIS monitor was observed when the BIS is not placed on the forehead, as seen by Suzuki et al. They used an atypical periauricular placement of the BIS monitor low on the patient’s face. Of note, this is not how the monitor is designed to be used. This method of positioning places the electrodes much further from the electrical impulses of the brain and nearer to the heart [6]. It is not surprising that in this location the cortical signal was weaker and the cardiac electrical signals much stronger. Electrical activity generated by an implanted cardiac pacemaker has also been shown to interfere with BIS monitors, resulting in erroneously high readings [7,8]. This may be due to the high electrical signal of the pacemaker compared to the intrinsic electrical activity of the heart. We report the appearance of cardiac electrical activity on a correctly placed BIS monitor in a neurologically intact patient under general anesthesia.

BIS monitoring is an important tool used by anesthesiologists to help titrate anesthetics to an appropriate depth. Even though this modality is generally reliable, anesthesia providers must be aware of potential sources of interference that can result in erroneous readings. The findings of this case report further highlight the notion that when monitoring the BIS as an indicator of the depth of anesthesia, much caution is needed. Many external sources of electrical “noise” may be interpreted as cortical “signals” and confuse the monitor. Misinterpreting these signals may lead to underestimating the depth of anesthesia. The subsequent overdosing of anesthetics may lead to patient harm. In this specific case, the patient was deeply anesthetized. This may have contributed to low EEG amplitudes that allowed for the relatively low energy signals of the patient’s cardiac cycle to be picked up and interpreted as true EEG signals. During surgery, close attention to the EEG waveform is necessary, as simply reading the number may lead to erroneous conclusions about the patient’s actual state of consciousness.

Anesthesiology, University of North Carolina at Chapel Hill School of Medicine, Chapel Hill, USA

Medicine, University of North Carolina at Chapel Hill, Chapel Hill, USA

Anesthesiology, University of North Carolina, Chapel Hill, USA

Anesthesiology, University of North Carolina at Chapel Hill School of Medicine, Chapel Hill, USA

Anesthesiology, University of North Carolina School of Medicine, Chapel Hill, USA

Anesthesiology, University of North Carolina at Chapel Hill School of Medicine, Chapel Hill, USA

Human subjects: All authors have confirmed that this study did not involve human participants or tissue. Conflicts of interest: In compliance with the ICMJE uniform disclosure form, all authors declare the following: Payment/services info: All authors have declared that no financial support was received from any organization for the submitted work. Financial relationships: All authors have declared that they have no financial relationships at present or within the previous three years with any organizations that might have an interest in the submitted work. Other relationships: All authors have declared that there are no other relationships or activities that could appear to have influenced the submitted work.

Doyal A S, Doyal H E, Flynn D N, et al. (September 23, 2022) Electrocardiographic Interference on Bispectral Index Monitor: A Case of Crossed Wires. Cureus 14(9): e29511. doi:10.7759/cureus.29511

Peer review began: September 14, 2022 Peer review concluded: September 19, 2022 Published: September 23, 2022

© Copyright 2022 Doyal et al. This is an open access article distributed under the terms of the Creative Commons Attribution License CC-BY 4.0., which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

Scholarly Impact Quotient™ (SIQ™) is our unique post-publication peer review rating process. Learn more here.

This link will take you to a third party website that is not affiliated with Cureus, Inc. Please note that Cureus is not responsible for any content or activities contained within our partner or affiliate websites.

Scholarly Impact Quotient™ (SIQ™) is our unique post-publication peer review rating process. SIQ™ assesses article importance and quality by embracing the collective intelligence of the Cureus community-at-large. All registered users are invited to contribute to the SIQ™ of any published article. (Authors cannot rate their own articles.)

High ratings should be reserved for work that is truly groundbreaking in its respective field. Anything above 5 should be considered above average. While all registered Cureus users can rate any published article, the opinion of domain experts is weighted appreciably more than that of non-specialists. An article’s SIQ™ will appear alongside the article after being rated twice and is recalculated with each additional rating.

Visit our SIQ™ page to find out more.

Scholarly Impact Quotient™ (SIQ™) is our unique post-publication peer review rating process. SIQ™ assesses article importance and quality by embracing the collective intelligence of the Cureus community-at-large. All registered users are invited to contribute to the SIQ™ of any published article. (Authors cannot rate their own articles.)

Already have an account? Sign in.

Please note that by doing so you agree to be added to our monthly email newsletter distribution list.

By joining Cureus, you agree to our Privacy Policy and Terms of Use.