TH1SCA-M is a hand-held informatization monitor that combines anesthetic gas analysis, end-tidal carbon dioxide (EtCO2) and oxygen saturation (SpO2) monitoring.
Different from traditional desktop monitors and anesthesia workstations, TH1SCA-M is more adaptable to a variety of complex environments, parameter monitoring is more intuitive, more convenient for a variety of occasions anesthetic gas analysis and ETC2 monitoring,
Gas Specifications
Different from traditional desktop monitors and anesthesia workstations, TH1SCA-M is more adaptable to a variety of complex environments, parameter monitoring is more intuitive, more convenient for a variety of occasions anesthetic gas analysis and ETC2 monitoring,
Gas Specifications
Sensor Type | Main stream | |
Principle | NDIR | |
Initialization time | Display the waveform within 3 seconds and meet the target requirements within 2 minutes (@ room temperature at 25ºC) | |
Compensations | Automatic atmospheric pressure compensation Optional gas compensation: oxygen (O2), laughing gas (N2O) compensation | |
Calibration | Zero point calibration is required after changing the adapter. | |
Measurement Range & Accuracy | ||
Gas | Measurement Range | Accuracy |
CO2 | 0-15 vol% | ±(0.2 vol% + 8% of reading) |
N2O | 0-100 vol% | ±(2 vol% + 2% of reading) |
SEV | 0-10 vol% | ±(0.15 vol% + 5% of reading) |
ISO | 0-8 vol% | ±(0.15 vol% + 5% of reading |
ENF | 0-8 vol% | ±(0.15 vol% + 5% of reading) |
Breath Rate | 3 - 150 bpm | ±1 bpm |
ResponseTime | < 0.5s | |
Breath detection algorithm | Adaptive threshold method | |
Waterproof | IPX4 - Splashproof (only the sensor itself) | |
Dimensions | 48mm X 24.3mm X 32.8mm |